Controls
For spectral cytometry,
we need to extract the spectra of the fluorochromes used one by
one.
To this end, it will be
necessary to provide:
1) Completely unstained
cells. The autofluorescence of cells is indeed considered as a
fluorochrome. It is therefore possible to evaluate the
autofluorescence spectrum and subtract it from the fluorescence
signals.
2) All single stained
(tubes containing cells stained with only one fluorochrome. The
preparation of single stained cells must be the same of the
sample (dissociation, washing, permeabilization, fixation,
etc.). If you are not sure about the presence of a marker on
your cells (or you have not enough cells for all controls) and
in order to extract the spectrum, you will need to provide beads
that bind your antibodies. The beads must undergo the same
treatments as that of cells. You can also find beads for
viability markers.
Be careful : Avoid using beads for
tandem fluorochromes (for instance : PE-Cy5.5, APC-Cy7, BV570,
QDOT655..) because the extraction
of the spectra is not done correctly when applied to cells :
Be careful : For unlabeled control we need minimum 100 000
cells (in about 200-300 µl) and for positive controls we need
minimum 200 positive cells in order to extract the spectra.
How to choose fluorochromes
In order to help you
in your choice of fluorochromes, Cytek has prepared a
downloadable document specific to our configuration:
https://welcome.cytekbio.com/hubfs/Website%20Downloadable%20Content/Data%20Sheets/Fluorochrome%20Guides/N9_20018_Rev._A_4L_VBYGR_Fluor_Guide.pdf
Also, do not forget the
rules for preparing a multimarker:
-the excitation wavelengths of the fluorochromes correspond
to the light sources of the cytometer,
-the emission wavelengths are far enough apart that their
signals can be analyzed separately.
-lowly expressed antigens are revealed by high yield
fluorochromes and highly expressed antigens with low yield
fluorochromes. To help you on this
point, the main fluorochromes are presented with their
brightness (stain index) in the following figure:
-in the case of
co-expression on a cell, use fluorochromes with little or no
overlapping spectra.
Elimination
of dead cells
The elimination of dead
cells can be done on the basis of the physical parameters side
scatter (SSC) and forward scatter (FSC) for cell line samples.
The addition of a dead cell marker (such as propidium iodide,
DAPI, Sytox, etc.) allows more precise elimination for more
complex samples. If a viability marker is used, the negative
control (unstained) must not contain the marker.
Cell filtration
Cell filtration, just
before analysis by the cell sorter, allows avoiding blockings
that destabilize cell sorting and decrease sorted fractions
purity. To this you can use 40 micron filters.
Conditioning cells
before sorting
The cells can be brought
in PBS or culture medium. The cells must be brought in 5ml tubes
(reference Falcon 2054) with a minimum volume of 300 µl at:
• 5x106 cells/ml maximum for low pressure sorting
• 1x107 cells/ml maximum for intermediate pressure sorting
• 2 to 5x107 cells/ml maximum for high pressure sorting
Be careful :The
larger the volume to pass, the longer the sort will take.
Collection tubes
The cells can be
collected after sorting in tubes (1.5ml Eppendorf tubes, 5ml
FACS tubes, 15ml Falcon tubes) or plates (96, 24,12 and 6
wells). It is advisable to put culture medium in the collection
tubes.
With the 100µ nozzle we
can get about :
• 1x106 cells in a 5ml tube
• 3x106 cells in a 15ml tube
with the 70µ
nozzle :
• 3x106 cells in a 5ml tube
• 9x106 cells in a 15ml tube
Adjust collect tubes
number to the collected cells estimated number for each
fraction.