Sorting form

https://www.mri.cnrs.fr/images/documents/E_MRI_DOC_0032_F_SORT_FORM.docx

It must be sent to us by email at the moment of booking so that we can validate the feasibility of  cell sorting (fluorochromes used and duration). For sorting with uncertain details at the moment of booking, the sorting form must be sent to us at the latest the day before the sorting. When you do the same experiment with the same markers several times, you can send a single sorting form indicating the time period concerned.

Number of “sortable” cells and sort duration
Two possible nozzle sizes (70 and 100 microns), which allow sorting a wide range of cell sizes. Enlarging the nozzle size also allows cells to be sorted at lower pressure (for fragile cells).
-At low pressure (100µ/15 PSI nozzle) the maximum speed is around 3 000 cells/second. It is therefore possible to sort a maximum of 11 000 000 cells/hour.
-At intermediate pressure (100µ/25 PSI nozzle) the maximum speed is around 5 000 cells/second. It is therefore possible to sort a maximum of 18 000 000 cells/hour.
-At high pressure (70µ/68 PSI nozzle) the maximum speed is around 20 000 cells/second. It is therefore possible to sort a maximum of 72 000 000 cells/hour.
Attention: These numbers take into account all events detected by the cell sorter including dead cells and debris. Take this into account in your calculations.
Calculation example: Although the sorting efficiency is often between 75% and 95%, it is always best to calculate the number of starting cells based on 50% yield rate so you wil be sure to have enough cells . If the population of interest represents for example 10% of your starting sample, to have 1x106 sorted cells it will be necessary to start from 2 x107 cells.


Example to calculate sorting duration
Nozzle size/Pressure

 100µ/15 PSI 100µ/25 PSI 70µ/68 PSI
Cell type  Fragile cells Intermediate conditions Small and not fragile cells
Average speed 1 000-3 000 evts/sec 2 000-5 000 evts/sec 10 000-20 000 evts/sec
Time needed* around 3 hours  around 1 hour and an half  around 20 minutes

*to sort 1M of cells of a population which represents 10 % of starting sample


Controls
For spectral cytometry, we need to extract the spectra of the fluorochromes used one by one.
To this end, it will be necessary to provide:
1) Completely unstained cells. The autofluorescence of cells is indeed considered as a fluorochrome. It is therefore possible to evaluate the autofluorescence spectrum and subtract it from the fluorescence signals.
2) All single stained (tubes containing cells stained with only one fluorochrome. The preparation of single stained cells must be the same of the sample (dissociation, washing, permeabilization, fixation, etc.). If you are not sure about the presence of a marker on your cells (or you have not enough cells for all controls) and in order to extract the spectrum, you will need to provide beads that bind your antibodies. The beads must undergo the same treatments as that of cells. You can also find beads for viability markers.
Be careful : Avoid using beads for tandem fluorochromes (for instance : PE-Cy5.5, APC-Cy7, BV570, QDOT655..) because the extraction of the spectra is not done correctly when applied to cells :




Be careful :
For unlabeled control we need minimum 100 000 cells (in about 200-300 µl) and for positive controls we need minimum 200 positive cells in order to extract the spectra.


How to choose fluorochromes
In order to help you in your choice of fluorochromes, Cytek has prepared a downloadable document specific to our configuration:
https://welcome.cytekbio.com/hubfs/Website%20Downloadable%20Content/Data%20Sheets/Fluorochrome%20Guides/N9_20018_Rev._A_4L_VBYGR_Fluor_Guide.pdf

Also, do not forget the rules for preparing a multimarker:
-the excitation wavelengths of the fluorochromes correspond to the light sources of the cytometer,
-the emission wavelengths are far enough apart that their signals can be analyzed separately.
-lowly expressed antigens are revealed by high yield fluorochromes and highly expressed antigens with low yield fluorochromes.
To help you on this point, the main fluorochromes are presented with their brightness (stain index) in the following figure:

 
Brillance
 -in the case of co-expression on a cell, use fluorochromes with little or no overlapping spectra.

Elimination of dead cells
The elimination of dead cells can be done on the basis of the physical parameters side scatter (SSC) and forward scatter (FSC) for cell line samples. The addition of a dead cell marker (such as propidium iodide, DAPI, Sytox, etc.) allows more precise elimination for more complex samples. If a viability marker is used, the negative control (unstained) must not contain the marker.

Cell filtration
Cell filtration, just before analysis by the cell sorter, allows avoiding blockings that destabilize cell sorting and decrease sorted fractions purity. To this  you can use 40 micron filters.

Conditioning cells before sorting
The cells can be brought in PBS or culture medium. The cells must be brought in 5ml tubes (reference Falcon 2054) with a minimum volume of 300 µl at:
     • 5x106 cells/ml maximum for low pressure sorting
     • 1x107 cells/ml maximum for intermediate pressure sorting
     • 2 to 5x107 cells/ml maximum for high pressure sorting
Be careful :The larger the volume to pass, the longer the sort will take.

Collection tubes
The cells can be collected after sorting in tubes (1.5ml Eppendorf tubes, 5ml FACS tubes, 15ml Falcon tubes) or plates (96, 24,12 and 6 wells). It is advisable to put culture medium in the collection tubes.
With the 100µ nozzle we can get about :
     • 1x106 cells in a 5ml tube
     • 3x106 cells in a 15ml tube
with the 70µ nozzle :
     • 3x106 cells in a 5ml tube
     • 9x106 cells in a 15ml tube
Adjust collect tubes number to the collected cells estimated number for each fraction.